confocal microscopy
scaling along z dimension,
u = { {2*\pi} \over {n \lambda}} NA^2 \times z
scaling transverse r dimension, assuming cylindrical symmetry
v = { {2*\pi} \over \lambda } NA \times r
u, v are called optical units. Born and Wolf, chapter 8.8
--
CM is one of the means by which we increase contrast so that the image more faithfully exhibits the available resolution. a CM does have slightly higher resolution than a wide field microscope, but this is not the source of most of its success.
--
PSF_{transverse} = 2J_1^2(v)/v^2 for a circular aperture. This is the familiar Airy disc. The pattern is characterized by the wavelength of light illuminating the circular aperture, and the aperture's size.
PSF_{axial} = ({ \sin{(u/4)} \over {u/4}})^2 for a paraxial approximation. This is a diffraction pattern of a slit.